Applied Mathematics and Mechanics (English Edition) ›› 1995, Vol. 16 ›› Issue (5): 437-442.

• Articles • Previous Articles     Next Articles

TRANSIENT ANALYSIS OF ARTIFICIAL MECHANICAL VALVE--BLOOD INTERACTION

Chen Da-peng1, Zhang Jian-hai1, Zou Sheng-quan2   

  1. 1. Institute of Computational Engineering Science, Southwest Jiaotong University, Chengdu 610031, P. R. China;
    2. Department of Engineering Mechanics, Sichuan Union University, Chengdu 610065, P. R. China
  • Received:1994-05-05 Online:1995-05-18 Published:1995-05-18
  • Supported by:

    Project supported by the National Natural Science Foundation and Doctoral Foundation of China

Abstract: Using finite elemcnt method,this paper has analyzed the blood-mechanical heart valve interaction system subjected to a step pressure when the vaiue is at closing position.As demonstrated in the present stud),,in such conditions mechanical values made of pyrolytic carbon,Ti alloy,Co-Cr alloy and ceramics tend to be very stiff which result in high impinging pressure.The impinging pressure acted on the value of the blood-valve sytem can be reduced by decreasing the elastic modulus of the mechanical value.

Key words: Stokes flow, Stokeslet, Fouries transform, harmonic function, biharmonic function, artificial mechanical valve, fluid-solid interaction, transient dynamic analysis, finite element method